CALIBRATION OF GAFCHROMIC EBT3 FILM FOR DOSIMETRY OF SCANNING PROTON PENCIL BEAM (PBS)
Stockholm University, Faculty of Science, Department of Physics.
Show others and affiliations
Number of Authors: 72018 (English)In: Radiation Protection Dosimetry, ISSN 0144-8420, E-ISSN 1742-3406, Vol. 180, no 1-4, p. 324-328Article in journal (Refereed) Published
Abstract [en]

Gafchromic EBT3 films are applied in proton radiotherapy for 2D dose mapping because they demonstrate spatial resolution well below 1mm. However, the film response must be corrected in order to reach the accuracy of dose measurements required for the clinical use. The in-house developed AnalyseGafchromic software allows to analyze and correct the measured response using triple channel dose calibration, statistical scan-to-scan fluctuations as well as experimentally determined dose and LET dependence. Finally, the optimized protocol for evaluation of response of Gafchromic EBT3 films was applied to determine 30 x 40 cm(2) dose profiles of the scanning therapy unit at the Cyclotron Centre Bronowice, CCB in Krakow, Poland.

Place, publisher, year, edition, pages
2018. Vol. 180, no 1-4, p. 324-328
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
URN: urn:nbn:se:su:diva-160261DOI: 10.1093/rpd/ncx304ISI: 000440983000068PubMedID: 29351653OAI: oai:DiVA.org:su-160261DiVA, id: diva2:1249428
Available from: 2018-09-19 Created: 2018-09-19 Last updated: 2018-09-19Bibliographically approved

Open Access in DiVA

No full text in DiVA

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Bassler, Niels
By organisation
Department of Physics
In the same journal
Radiation Protection Dosimetry
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 11 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf