Commissioning and validation of small subfields in Step-and-shoot IMRT
Stockholm University, Faculty of Science, Medical Radiation Physics (together with KI).
2008 (English)Independent thesis Basic level (professional degree), 20 credits / 30 HE creditsStudent thesis
Abstract [en]

One of the most used irradiation techniques in modern radiation therapy is step-and-shoot IMRT. The accuracy of this technique when delivering complex dose distributions strongly depends on the size of the subfields. The aims of this study is to determine the minimum size of subfields that can be used efficiently in Step-and-Shoot IMRT, to investigate the validation process for beam delivery and treatment planning dose calculations, and to find recommendations for practical clinical implementations.

Two different detectors, a CC04 ion chamber and a SFD stereotactic diode, have been used for measuring head scatter factors in air (Sc), total output factors (Scp) and dose profiles in water for a wide range of field sizes. The measurements were compared to calculations done with a pre-release version of the Nucletron MasterPlanTM v 3.1 treatment planning system that employs a novel, high resolution fluence modelling for both its pencil beam and collapsed cone dose calculation algorithms. Collimator settings were explicitly checked using FWHM film measurements with a build-up sheet of tungsten placed close to the treatment head to reduce the influence from lateral electron transport and geometrical penumbra. An analysis of the influence and sensitivity of Scp for small fields with respect to the linear accelerator source size and shape was also made.

The measurements with the ionization chamber and the stereotactic diode showed good agreements with each other and with the treatment planning system calculations for field sizes larger than 2×2 cm2. For small field sizes, measurements with different detectors yielded different results. Calculations showed agreements with measurements with the smallest detector, provided careful field size calibration and commissioning of calculation parameters. Uncertainties in collimator settings and source characteristics were shown to yield large uncertainties in Scp for fields smaller than 2×2 cm2.

The treatment planning system was found to properly handle small subfields but results were very sensitive to uncertainties in source size, as well as calibration and reproducibility of the collimator settings. Therefore if subfields smaller than 2×2 cm2 are to be used in IMRT extra care should be taken to determine the source characteristics and to calibrate the collimators. The volume of the detectors used for validation of such small fields and the loss of charged particle equilibrium conditions also have to be taken into consideration.

Place, publisher, year, edition, pages
2008. , 68 p.
Keyword [en]
Radiotherapy Planning, Radiotherapy, Intensity-Modulated, IMRT, Commissioning, Small fields
National Category
Radiology, Nuclear Medicine and Medical Imaging
URN: urn:nbn:se:su:diva-8316OAI: diva2:200074
Available from: 2008-11-06 Created: 2008-11-06Bibliographically approved

Open Access in DiVA

fulltext(2270 kB)1877 downloads
File information
File name FULLTEXT01.pdfFile size 2270 kBChecksum MD5
Type fulltextMimetype application/pdf

By organisation
Medical Radiation Physics (together with KI)
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 1877 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Total: 458 hits
ReferencesLink to record
Permanent link

Direct link