Dosimetric Characteristics of CVD Single Crystal Diamond Detectors in Radiotherapy Beams
Stockholm University, Faculty of Science, Medical Radiation Physics (together with KI).
2008 (English)Independent thesis Basic level (professional degree), 20 credits / 30 HE creditsStudent thesis
Abstract [en]

Dosimetric characteristics of a CVD single crystal diamond detector have been evaluated. Detector stability, linearity, optimal bias, temperature dependence, directional dependence, priming and pre-irradiation behaviour, depth dose curves and dose profiles were investigated.

The optimal bias was determined to be 50 V. The detector stability measurement showed a too large variation for absolute dosimetry in a day to day measurement, but acceptable variation during one and the same day. The linearity constant, , in the relation between signal and dose rate, (Fowler 1966), was determined to 0.978 and 0.953 for two detectors. The sub-linearity was also observed in the depth dose curves and could be eliminated with a correction method. The diamond detector showed smaller temperature dependence than the EFD silicon diode. The directional dependency was, <1 %, up to at least ± 15˚ and therefore no angular correction is needed. A priming dose of 0.6 Gy was determined, which is considerably smaller than for existing detectors on the market. After pre-irradiation with electrons (8 and 18 MeV) a large and permanent desensitization of up to 31 % / 500 Gy was detected. This is in contradiction to what previous published articles claim. 15 MV photons also reduced the sensitivity of the detector, but no evidence that 5 MV photons do has been found. A 50 Gy dose of 180 MeV protons did not reduce the sensitivity either. The detector dose rate linearity was improved by electron pre-irradiation. The dose profile penumbras of the diamond detector were, for the most part, smaller than the RK ionization chamber, indicating a better spatial resolution.

Place, publisher, year, edition, pages
2008. , 48 p.
Keyword [en]
Diamond detector, CVD, Priming dose
National Category
Radiology, Nuclear Medicine and Medical Imaging
URN: urn:nbn:se:su:diva-8321OAI: diva2:200080
Available from: 2008-11-06 Created: 2008-11-06Bibliographically approved

Open Access in DiVA

fulltext(243 kB)555 downloads
File information
File name FULLTEXT01.pdfFile size 243 kBChecksum MD5
Type fulltextMimetype application/pdf

By organisation
Medical Radiation Physics (together with KI)
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
Total: 555 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Total: 224 hits
ReferencesLink to record
Permanent link

Direct link