Influence of phantom thickness and material on the backscatter factors for diagnostic x-ray beam dosimetry
Stockholm University, Faculty of Science, Department of Physics. Karolinska Institutet, Sweden.
Stockholm University, Faculty of Science, Department of Physics. Karolinska Institutet, Sweden.
2013 (English)In: Physics in Medicine and Biology, ISSN 0031-9155, E-ISSN 1361-6560, Vol. 58, no 2, 247-260 p.Article in journal (Refereed) Published
Abstract [en]

Most of the existing backscatter factors for the dosimetry of clinical diagnostic x-ray beams have been calculated for 15 cm thick phantoms; these data are used for skin dose determinations which in general ignore the influence of phantom material and thickness. The former should strictly be required whenever dosimetry measurements are made on phantom materials different from those used for the backscatter factor calculations. The phantom or patient thickness is of special importance when skin dose determinations are made for infants or paediatric patients. In this work, the recently published formalism for reference dosimetry and comprehensive database of backscatter factors for clinical beams and water phantoms have been extended using two correction factors which account for phantom material and thickness. These were determined with simulations using the PENELOPE Monte Carlo system, for PMMA to analyse the influence of the phantom material relative to water, and for a broad range of thicknesses of water and PMMA to investigate the role of this parameter in patient dose estimates. The material correction factor was found to be in the range 3-10%, depending on the field size and the HVL. The thickness correction factor was in the range 2-12% for a 5 cm thick phantom and square field sizes between 5 and 35 cm, reaching a plateau of about ±1% for thicknesses beyond 13 cm. Expressions in the form of surface fits over the calculated data are provided which streamline the determination of backscatter factors for arbitrary thicknesses and phantom materials, as well as field sizes. Results demonstrate the inadequacy of using conventional backscatter factors (calculated for 15 cm thick phantoms) without correction factors that take into account the phantom material and its thickness.

Place, publisher, year, edition, pages
2013. Vol. 58, no 2, 247-260 p.
Keyword [en]
small field dosimetry, Monte Carlo, PENELOPE, output correction factors
National Category
Physical Sciences
Research subject
Medical Radiation Physics
Identifiers
URN: urn:nbn:se:su:diva-114133DOI: 10.1088/0031-9155/58/2/247ISI: 000312674400006PubMedID: 23257169OAI: oai:DiVA.org:su-114133DiVA: diva2:789934
Available from: 2015-02-20 Created: 2015-02-20 Last updated: 2017-12-04Bibliographically approved
In thesis
1. Key Data for the Reference and Relative Dosimetry of Radiotherapy and Diagnostic and Interventional Radiology Beams
Open this publication in new window or tab >>Key Data for the Reference and Relative Dosimetry of Radiotherapy and Diagnostic and Interventional Radiology Beams
2015 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Accurate dosimetry is a fundamental requirement for the safe and efficient use of radiation in medical applications. International Codes of Practice, such as IAEA TRS-398 (2000) for radiotherapy beams and IAEA TRS-457 (2007) for diagnostic radiology beams, provide the necessary formulation for reference and relative dosimetry and the data required for their implementation. Research in recent years has highlighted the shortage of such data for radiotherapy small photon beams and for surface dose estimations in diagnostic and interventional radiology, leading to significant dosimetric errors that in some instances have jeopardized patient’s safety and treatment efficiency.

The aim of this thesis is to investigate and determine key data for the reference and relative dosimetry of radiotherapy and radiodiagnostics beams. For that purpose the Monte Carlo system PENELOPE has been used to simulate the transport of radiation in different media and a number of experimental determinations have also been made. A review of the key data for radiotherapy beams published after the release of IAEA TRS-398 was conducted, and in some cases the considerable differences found were questioned under the criterion of data consistency throughout the dosimetry chain (from standards laboratories to the user). A modified concept of output factor, defined in a new international formalism for the dosimetry of small photon beams, requires corrections to dosimeter readings for the dose determination in small beams used clinically. In this work, output correction factors were determined, for Varian Clinac 6 MV photon beams and Leksell Gamma Knife Perfexion 60Co gamma-ray beams, for a large number of small field detectors, including air and liquid ionization chambers, shielded and unshielded silicon diodes and diamond detectors, all of which were simulated by Monte Carlo with great detail.

Backscatter factors and ratios of mass energy-absorption coefficients required for surface (skin) determinations in diagnostic and interventional radiology applications were also determined, as well as their extension to account for non-standard phantom thicknesses and materials. A database of these quantities was created for a broad range of monoenergetic photon beams and computer codes developed to convolve the data with clinical spectra, thus enabling the determination of key data for arbitrary beam qualities.

Data presented in this thesis has been contributed to the IAEA international dosimetry recommendations for small radiotherapy beams and for diagnostic radiology in paediatric patients.

Place, publisher, year, edition, pages
Stockholm: Department of Physics, Stockholm University, 2015. 68 p.
Keyword
Backscatter factors, Diagnostic radiology dosimetry, Mass energy-absorption coefficients, Monte Carlo, Output correction factors, Radiotherapy dosimetry, Reference dosimetry, Relative dosimetry, Small photon fields
National Category
Physical Sciences
Research subject
Medical Radiation Physics
Identifiers
urn:nbn:se:su:diva-114413 (URN)978-91-7649-111-9 (ISBN)
Public defence
2015-04-22, Föreläsningssalen, (P1:01), Radiumhemmet, Karolinska Universitetssjukhuset, Solna, 14:30 (English)
Opponent
Supervisors
Note

At the time of the doctoral defense, the following paper was unpublished and had a status as follows: Paper 6: Manuscript.

Available from: 2015-03-31 Created: 2015-03-03 Last updated: 2015-05-12Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full textPubMed

Search in DiVA

By author/editor
Benmakhlouf, HamzaAndreo, Pedro
By organisation
Department of Physics
In the same journal
Physics in Medicine and Biology
Physical Sciences

Search outside of DiVA

GoogleGoogle Scholar

doi
pubmed
urn-nbn

Altmetric score

doi
pubmed
urn-nbn
Total: 132 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf