Technical Note: On the calculation of stopping-power ratio for stoichiometric calibration in proton therapy
Stockholm University, Faculty of Science, Department of Physics. Karolinska Institutet, Sweden.
Show others and affiliations
Number of Authors: 5
2015 (English)In: Medical physics (Lancaster), ISSN 0094-2405, Vol. 42, no 9, 5252-5257 p.Article in journal (Refereed) Published
Abstract [en]

Purpose: The quantitative effects of assumptions made in the calculation of stopping-power ratios (SPRs) are investigated, for stoichiometric CT calibration in proton therapy. The assumptions investigated include the use of the Bethe formula without correction terms, Bragg additivity, the choice of I-value for water, and the data source for elemental I-values. Methods: The predictions of the Bethe formula for SPR (no correction terms) were validated against more sophisticated calculations using the SRIM software package for 72 human tissues. A stoichiometric calibration was then performed at our hospital. SPR was calculated for the human tissues using either the assumption of simple Bragg additivity or the Seltzer-Berger rule (as used in ICRU Reports 37 and 49). In each case, the calculation was performed twice: First, by assuming the I-value of water was an experimentally based value of 78 eV (value proposed in Errata and Addenda for ICRU Report 73) and second, by recalculating the I-value theoretically. The discrepancy between predictions using ICRU elemental I-values and the commonly used tables of Janni was also investigated. Results: Errors due to neglecting the correction terms to the Bethe formula were calculated at less than 0.1% for biological tissues. Discrepancies greater than 1%, however, were estimated due to departures from simple Bragg additivity when a fixed I-value for water was imposed. When the I-value for water was calculated in a consistent manner to that for tissue, this disagreement was substantially reduced. The difference between SPR predictions when using Janni's or ICRU tables for I-values was up to 1.6%. Experimental data used for materials of relevance to proton therapy suggest that the ICRU-derived values provide somewhat more accurate results (root-mean-square-error: 0.8% versus 1.6%). Conclusions: The conclusions from this study are that (1) the Bethe formula can be safely used for SPR calculations without correction terms; (2) simple Bragg additivity can be reasonably assumed for compound materials; (3) if simple Bragg additivity is assumed, then the I-value for water should be calculated in a consistent manner to that of the tissue of interest (rather than using an experimentally derived value); (4) the ICRU Report 37 I-values may provide a better agreement with experiment than Janni's tables. (C) 2015 American Association of Physicists in Medicine.

Place, publisher, year, edition, pages
2015. Vol. 42, no 9, 5252-5257 p.
Keyword [en]
Proton therapy, stopping-power, CT calibration, stoichiometric method, I-values
National Category
Radiology, Nuclear Medicine and Medical Imaging
URN: urn:nbn:se:su:diva-121493DOI: 10.1118/1.4928399ISI: 000360645000025OAI: diva2:860606
Available from: 2015-10-13 Created: 2015-10-05 Last updated: 2015-10-13Bibliographically approved

Open Access in DiVA

No full text

Other links

Publisher's full text

Search in DiVA

By author/editor
Ödén, Jakob
By organisation
Department of Physics
In the same journal
Medical physics (Lancaster)
Radiology, Nuclear Medicine and Medical Imaging

Search outside of DiVA

GoogleGoogle Scholar
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

Altmetric score

Total: 13 hits
ReferencesLink to record
Permanent link

Direct link