Drawbacks of the current NCRP models on shielding calculations: Development of a new methodology for the Gamma Knife®
Stockholms universitet, Naturvetenskapliga fakulteten, Fysikum. Eduardo Mondlane University, Mozambique.
Stockholms universitet, Naturvetenskapliga fakulteten, Fysikum.
2022 (Engelska)Konferensbidrag, Poster (med eller utan abstract) (Övrigt vetenskapligt)
Abstract [en]

Most practical guidelines for design of shielding against photon radiotherapy equipment may be retrieved from the National Council on Radiation Protection and Measurements (NCRP) report 151. The NCRP report No. 151 formalism describes structural shielding design techniques for megavoltage x- and gamma-ray radiotherapy facilities with the aim to limit radiation exposures to members of the public, employees, and the environment to an acceptable level. The implemetaion of NCRP report No. 151 recommendations may be straightforward for linear accelerators and conventional Co-60 machines which have singular source of radiation in contrast to the Leksell Gamma Knife, which has a complex multisource configuration. To harness the full potential of the NCRP 151 methodology, complementary transmission measurements may be required. In the NCRP report No. 151 the primary barriers are designed to attenuate photons emanating directly from the treatment unit while secondary barriers are calculated to shield radiation leakage and scattered photons by the patient and objects inside the treatment room. Here it is assumed that a primary radiation barrier should at the same time shield against secondary radiation. In this work the gamma-ray spectra in the vicinity of the Elekta Leksell Gamma Knife-Perfexion were acquired using a High Purity Germanium detector. It has been observed that the proportion of high energetic photons is lower in comparison with photons with low energy. This corroborates with the fact that the Leksell Gamma Knife has been designed in order to prevent any primary radiation from exiting the unit without impinging on an internal interceptor shield during all states of the machine. Therefore, the need to adapt the NCRP methodology for the Gamma Knife appeared to be justified.

Ort, förlag, år, upplaga, sidor
2022.
Nyckelord [en]
Gamma knife Perfexion, HPGe detector, gamma-ray spectroscopy, NCRP
Nationell ämneskategori
Subatomär fysik Radiologi och bildbehandling
Identifikatorer
URN: urn:nbn:se:su:diva-233383OAI: oai:DiVA.org:su-233383DiVA, id: diva2:1896739
Konferens
6th International Symposium on the System of Radiological Protection (ICRP 2021+1), Vancouver, Canada, 7-10 November, 2022
Tillgänglig från: 2024-09-11 Skapad: 2024-09-11 Senast uppdaterad: 2025-02-14Bibliografiskt granskad
Ingår i avhandling
1. Optimization of brachytherapy treatment planning and radiation field studies in the vicinity of the Leksell Gamma Knife®
Öppna denna publikation i ny flik eller fönster >>Optimization of brachytherapy treatment planning and radiation field studies in the vicinity of the Leksell Gamma Knife®
2024 (Engelska)Doktorsavhandling, sammanläggning (Övrigt vetenskapligt)
Abstract [en]

Carcinoma of the cervix is a global problem, and brachytherapy is one of the main radiation therapy components used in the treatment of patients with this type of cancer. With the advent of scientific and technological developments in treatment planning, inverse optimization in brachytherapy and its thorough comparison with traditional manual optimization methods is warranted. In this work, the physical parameters; minimum dose received by 98% and 90% of the target volume represented by D98 and D90, respectively, were used to evaluate the treatment plans with respect to the target while the minimum dose received by 2cm3 volume, D2cm3, was used to investigate complications in organs at risk. The conformity index, COIN, was used to describe the coverage of the target by the prescribed dose and the fraction of each organ at risk volume that receives the critical dose, which may cause complications. The treatment plan evaluation was also performed in terms of the radiobiological parameter complication-free tumor control probability, P+. The physical and radiobiological evaluation corresponding to two brachytherapy inverse planning algorithms have been compared with homologous manual graphical optimization plans. The main observations from this study are that well tuned class solutions of inverse optimization methods may produce similar dose volume histograms to those produced with manual graphical optimization plans, and inverse methods have the potential to spare organs at risk while delivering an acceptable dose to the target. In addition, radiobiological indexes such as the P+ can be useful complements to the physical parameters in treatment plan evaluation. The Elekta Leksell Gamma Knife® unit has been successfully utilized in the management of intracranial malignancies for more than half a century. As required by national and international regulation, adequate knowledge of risks posed by ionizing radiation instrumentation is necessary in order to protect the patient, workers, public, and the environment. In that perspective, the Nuclear Physics research group at the Department of Physics of Stockholm University (Stockholm, Sweden) in collaboration with Elekta Instrument AB (Stockholm, Sweden) have conducted investigation on the radiation field in the vicinity of the Gamma Knife utilizing the High Purity Germanium (HPGe) detector. As part of the ongoing research, the principal objective of the present work was to improve the modeling and characterization of the radiation field around the Gamma Knife to interrogate the efficacy of the National Council on Radiation Protection and Measurements (NCRP) methodology for structural shielding design and evaluation for the Leksell Gamma Knife treatment room. Acquisition of high-resolution γ-ray spectra and ambient dose equivalent H*(10) in the field of the Gamma Knife-PerfexionTM took place at the Division of Neurosurgery, Radiumhemmet (Department of Oncology), Karolinska University Hospital (Stockholm, Sweden). A p-type coaxial HPGe detector and a satellite survey meter were utilized to acquire the γ-ray spectra and H*(10), respectively. The measured configurations were simulated on Pegasos Monte Carlo system. A phase space on a cylindrical surface enclosing the Gamma Knife with open doors and the phantom assembled was used as the source of radiation. About 4·107 γ photons were collected on the phase space corresponding to 2·1012 decays. With the Gamma Knife doors open, most of the radiation was measured in the forward direction up to θ = 45o in relation to the z axis. The Monte Carlo simulations reproduced the measured results and; therefore, good agreement was achieved between response measurements and simulated spectra. The recent Gamma Knife models PerfexionTM, IconTM, and EspritTM are dosimetricaly equivalent and represent more than half the total number of Gamma Knife systems installed around the globe. Given the increasing importance of these machines in the treatment of patients with intracranial malignancies, research and development of tools for radiological protection around the Gamma Knife unit is topical and will continue to evolve, especially with the advent of new models or versions of Gamma Knife units integrating novel technology such as the satellite cone-beam computed tomography (CBCT) imaging system. The Monte Carlo method is a powerful tool for simulation of radiation transport in matter. It reproduced the measured results and may be used in structural shielding design calculations for γ-photons from the Gamma Knife.

Ort, förlag, år, upplaga, sidor
Stockholm: Department of Physics, Stockholms universitet, 2024. s. 108
Nyckelord
Brachytherapy, treatment planning optimization, radiobiology, complication-free tumor control probability, Leksell Gamma Knife, stereotactic radiation surgery, HPGe detector, detector response, Monte Carlo simulation, dosimetry, radiological protection
Nationell ämneskategori
Fysik
Forskningsämne
medicinsk strålningsfysik
Identifikatorer
urn:nbn:se:su:diva-233393 (URN)978-91-8014-933-4 (ISBN)978-91-8014-934-1 (ISBN)
Disputation
2024-10-25, Room FB42, AlbaNova Campus, Roslagstullsbacken 21, Stockholm, 10:00 (Engelska)
Opponent
Handledare
Tillgänglig från: 2024-10-02 Skapad: 2024-09-11 Senast uppdaterad: 2024-10-08

Open Access i DiVA

Fulltext saknas i DiVA

Person

dos Santos Matias, LucílioTegnér, Per-Erik

Sök vidare i DiVA

Av författaren/redaktören
dos Santos Matias, LucílioTegnér, Per-Erik
Av organisationen
Fysikum
Subatomär fysikRadiologi och bildbehandling

Sök vidare utanför DiVA

GoogleGoogle Scholar

urn-nbn

Altmetricpoäng

urn-nbn
Totalt: 187 träffar
RefereraExporteraLänk till posten
Permanent länk

Direktlänk
Referera
Referensformat
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Annat format
Fler format
Språk
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Annat språk
Fler språk
Utmatningsformat
  • html
  • text
  • asciidoc
  • rtf