Robust optimization considering uncertainties in adaptive proton therapy.
Stockholm University, Faculty of Science, Department of Physics. RaySearch Laboratories AB and Karolinska Institutet, Stockholm, Sweden.ORCID iD: 0000-0002-8441-3595
2024 (English)Licentiate thesis, comprehensive summary (Other academic)
Abstract [en]

Proton therapy, a promising alternative to conventional photon therapy, has gained widespread acceptance in clinical practice. This is attributed to its superior depth-dose curve that has a negligible dose beyond the maximum range of the proton. A proton treatment planning requires a multitude of parameters and are either manually selected or optimized using mathematical formulation. However, a proton treatment plan is also subject to various systematic and random uncertainties that must be taken into account during optimization. Robust optimization is a commonly used method for integrating the setup and range uncertainties in proton therapy.

In addition to the uncertainties accounted for during the treatment planning phase, others can arise during the course of treatment and are often hard to predict. Changes in the patient's anatomy represent uncertainties that can significantly affect planned dose delivery. Therefore, adaptive planning is typically performed intermittently or regularly, depending on the changes in anatomy. Paper II included in this thesis proposed a method of adaptive planning that takes into account the impact of the patient's respiratory motion at the treatment site, such as the lungs and abdomen for 4D robust optimization. This method uses dose mimicking to reproduce the results as initially planned.  

This additional stage of adaptive planning can introduce new complexities and uncertainties into the treatment process. One such uncertainty arise from daily cone beam computed tomography (CBCT) images which are required for treatment plan adaptation. Several strategies have been proposed in the past to improve the quality of these images, but each strategy has its advantages and disadvantages, depending on the site of treatment. In Paper I, a method was proposed that combined the advantages of other frequently used methods to create an improved method for generating daily images with CT-like image quality. This can contribute towards the goal of online adaptive in the near future with reduced uncertainties.

This thesis will provide a brief introduction and an in-depth chapter to elucidate the background, better understand the physics of proton therapy, the process of treatment planning, and the need for adaptive planning.

Place, publisher, year, edition, pages
Stockholm: Department of Physics, Stockholm University , 2024.
Keywords [en]
Proton therapy, Robust optimization, Adaptive radiation therapy
National Category
Physical Sciences
Research subject
Medical Radiation Physics
Identifiers
URN: urn:nbn:se:su:diva-229129OAI: oai:DiVA.org:su-229129DiVA, id: diva2:1857431
Presentation
2024-06-03, CCK Lecture Hall, Cancer Center Karolinska (CCK), Karolinska Universitetssjukhuset, 171 76 Solna, Solna, 09:00 (English)
Opponent
Supervisors
Projects
European Union’s Horizon 2020 Marie Skłodowska-Curie Actions under Grant Agreement No. 955956
Funder
EU, Horizon 2020, 955956Available from: 2024-05-13 Created: 2024-05-13 Last updated: 2024-05-13Bibliographically approved
List of papers
1. Generation and evaluation of anatomy-preserving virtual CT for online adaptive proton therapy
Open this publication in new window or tab >>Generation and evaluation of anatomy-preserving virtual CT for online adaptive proton therapy
Show others...
2024 (English)In: Medical Physics, ISSN 0094-2405, E-ISSN 2473-4209, Vol. 51, no 3, p. 1536-1546Article in journal (Refereed) Published
Abstract [en]

Background: Daily CTs generated by CBCT correction are required for daily replanning in online-adaptive proton therapy (APT) to effectively deal with inter-fractional changes. Out of the currently available methods, the suitability of a daily CT generation method for proton dose calculation also depends on the anatomical site.

Purpose: We propose an anatomy-preserving virtual CT (APvCT) method as a hybrid method of CBCT correction, which is especially suitable for large anatomy deformations. The accuracy of the hybrid method was assessed by comparison with the corrected CBCT (cCBCT) and virtual CT (vCT) methods in the context of online APT.

Methods: Seventy-one daily CBCTs of four prostate cancer patients treated with intensity modulated proton therapy (IMPT) were converted to daily CTs using cCBCT, vCT, and the newly proposed APvCT method. In APvCT, planning CT (pCT) were mapped to CBCT geometry using deformable image registration with boundary conditions on controlling regions of interest (ROIs) created with deep learning segmentation on cCBCT. The relative frequency distribution (RFD) of HU, mass density and stopping power ratio (SPR) values were assessed and compared with the pCT. The ROIs in the APvCT and vCT were compared with cCBCT in terms of Dice similarity coefficient (DSC) and mean distance-to-agreement (mDTA). For each patient, a robustly optimized IMPT plan was created on the pCT and subsequent daily adaptive plans on daily CTs. For dose distribution comparison on the same anatomy, the daily adaptive plans on cCBCT and vCT were recalculated on the corresponding APvCT. The dose distributions were compared in terms of isodose volumes and 3D global gamma-index passing rate (GPR) at γ(2%, 2 mm) criterion.

Results: For all patients, no noticeable difference in RFDs was observed amongst APvCT, vCT, and pCT except in cCBCT, which showed a noticeable difference. The minimum DSC value was 0.96 and 0.39 for contours in APvCT and vCT respectively. The average value of mDTA for APvCT was 0.01 cm for clinical target volume and ≤0.01 cm for organs at risk, which increased to 0.18 cm and ≤0.52 cm for vCT. The mean GPR value was 90.9%, 64.5%, and 67.0% for APvCT versus cCBCT, vCT versus cCBCT, and APvCT versus vCT, respectively. When recalculated on APvCT, the adaptive cCBCT and vCT plans resulted in mean GPRs of 89.5 ± 5.1% and 65.9 ± 19.1%, respectively. The mean DSC values for 80.0%, 90.0%, 95.0%, 98.0%, and 100.0% isodose volumes were 0.97, 0.97, 0.97, 0.95, and 0.91 for recalculated cCBCT plans, and 0.89, 0.88, 0.87, 0.85, and 0.81 for recalculated vCT plans. Hausdorff distance for the 100.0% isodose volume in some cases of recalculated cCBCT plans on APvCT exceeded 1.00 cm.

Conclusions: APvCT contours showed good agreement with reference contours of cCBCT which indicates anatomy preservation in APvCT. A vCT with erroneous anatomy can result in an incorrect adaptive plan. Further, slightly lower values of GPR between the APvCT and cCBCT-based adaptive plans can be explained by the difference in the cCBCT's SPR RFD from the pCT.

Keywords
adaptive proton therapy, virtual/synthetic CT | daily anatomical variation
National Category
Radiology, Nuclear Medicine and Medical Imaging
Identifiers
urn:nbn:se:su:diva-226118 (URN)10.1002/mp.16941 (DOI)001143329900001 ()38230803 (PubMedID)2-s2.0-85182451194 (Scopus ID)
Available from: 2024-02-06 Created: 2024-02-06 Last updated: 2026-01-19
2. Adaptive intensity-modulated proton therapy with 4D robust planning: A dose mimicking approach
Open this publication in new window or tab >>Adaptive intensity-modulated proton therapy with 4D robust planning: A dose mimicking approach
Show others...
(English)Manuscript (preprint) (Other academic)
Abstract [en]

Objective: A 4D robust optimisation (4DRO) is usually employed when the tumour respiratory motion needs to be addressed. However, it is computationally demanding, and an automated method is preferable for adaptive planning to avoid manual trial-and-error. This study proposes a 4DRO technique based on dose mimicking for adaptive planning.

Approach: Initial plans for 4D robust IMPT were created on an average CT (avgCT) for four patients with clinical target volume (CTV) in the lung, oesophagus, or pancreas, respectively. These plans were robustly optimized using three phases of 4DCT and accounting for setup and density uncertainties. Weekly 4DCTs were used for adaptive replanning, using a constant relative biological effectiveness (cRBE) of 1.1. Two methods were used for this purpose: (1) template-based adaptive (TA) planning and (2) dose-mimicking-based adaptive (MA) planning. The plans were evaluated using variable RBE (vRBE) weighted doses and biologically consistent dose accumulation (BCDA).

Main results: TA and MA plans had comparable CTV coverage except for one case where MA plan had a higher D98 and D2 but with an increased D2 in an OAR. CTV D98 deviations in non-adaptive plans from the initial plans were up to -7.2% in individual plans and -1.8% when using BCDA. For the OARs, MA plans showed a reduced mean dose and D2 compared to the TA plan, with few exceptions. The vRBE-weighted dose had a mean dose and D2 difference of up to 0.3 Gy and 0.5 Gy, respectively, in the OARs with respect to cRBE-weighted dose.

Significance: MA plans indicate better performance in target coverage and OAR dose sparing compared to the TA plans. Moreover, MA method is capable of handling both forms of anatomical variation, namely, changes in density and relative shifts in the position of the OARs.

National Category
Cancer and Oncology Other Physics Topics
Research subject
Medical Radiation Physics
Identifiers
urn:nbn:se:su:diva-229127 (URN)
Projects
European Union’s Horizon 2020 Marie Skłodowska-Curie Actions under Grant Agreement No. 955956
Available from: 2024-05-13 Created: 2024-05-13 Last updated: 2024-05-13

Open Access in DiVA

fulltext(6511 kB)563 downloads
File information
File name FULLTEXT01.pdfFile size 6511 kBChecksum SHA-512
de5cc31840c515b8dcfc3a2e085767bf2b2da2106b97597cd9908ba685fb96f308705b7a580f473048be8a5da1fb55758065d34914993378defa87f31af9c745
Type fulltextMimetype application/pdf

Search in DiVA

By author/editor
Kaushik, Suryakant
By organisation
Department of Physics
Physical Sciences

Search outside of DiVA

GoogleGoogle Scholar
Total: 564 downloads
The number of downloads is the sum of all downloads of full texts. It may include eg previous versions that are now no longer available

urn-nbn

Altmetric score

urn-nbn
Total: 834 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf