Radiobiological treatment planning evaluation of inverse planning simulated annealing for cervical cancer high-dose-rate brachytherapy
Stockholm University, Faculty of Science, Department of Physics. Eduardo Mondlane University, Mozambique.
Show others and affiliations
2015 (English)In: Anticancer Research, ISSN 0250-7005, E-ISSN 1791-7530, Vol. 35, no 2, p. 935-939Article in journal (Refereed) Published
Abstract [en]

Aim: To compare five inverse treatment plans with one conventional manually optimized plan for cervical cancer brachytherapy (BT) using radiobiological parameters combined with dosimetric and volumetric parameters.

Materials and Methods: Five inverse treatment plans were calculated using an inverse planning simulated annealing (IPSA) algorithm for each of four fractions for 12 cervical cancer patients treated with high-dose-rate (HDR) brachytherapy. The inverse treatment plans were compared with a manually optimized plan used for the actual treatment of the patients. The comparison of the plans was performed with respect to the probability of cure without complication.

Results: Overall, the manually optimized plan scored the best results; however, the probability of cure without complication is within an acceptable clinical range for all the plans.

Conclusion: Although there are still considerable uncertainties in the radiobiological parameters, the radiobiological plan evaluation method presents itself as a potential complement to physical dosimetric methods.

Place, publisher, year, edition, pages
2015. Vol. 35, no 2, p. 935-939
Keywords [en]
Brachytherapy, IPSA, radiobiological evaluation
National Category
Cancer and Oncology
Identifiers
URN: urn:nbn:se:su:diva-109064ISI: 000349598600041PubMedID: 25667477Scopus ID: 2-s2.0-84923646701OAI: oai:DiVA.org:su-109064DiVA, id: diva2:762190
Available from: 2014-11-10 Created: 2014-11-10 Last updated: 2024-09-11Bibliographically approved
In thesis
1. Optimization of brachytherapy for cervical cancer using inverse planning algorithms.
Open this publication in new window or tab >>Optimization of brachytherapy for cervical cancer using inverse planning algorithms.
2015 (English)Licentiate thesis, comprehensive summary (Other academic)
Abstract [en]

Carcinoma of the cervix is a global problem. Brachytherapy (BT) is one of the main

radiation therapy components used in the management of cervical cancer. With the

advent of scientific and technological developments in treatment planning, inverse optimization

in BT has been imposed; however, to harness the full potential of inverse

planning in brachytherapy, its thorough comparison with manual optimization methods

is warranted.

Although inverse optimization algorithms are based on different mathematical approaches,

their goals are similar. The underlying principles of these algorithms will

allow them to be applied with the aim of respecting normal structures absorbed dose

limits while delivering high enough tumouricidal dose.

In this work, the physical parameters minimum dose received by 98% and 90% of

the target volume represented by D98 and D90, respectively, were used to evaluate the

treatment plans with respect to the target while the minimum dose received by 2cm3

volume, D2cm3 , was used to investigate complications in organs at risk (OARs). The

conformity index (COIN), was used to describe the coverage of the target by the prescribed

dose (PD) and the fraction of each OAR volume that receives a critical dose,

which may cause complication. The treatment plan evaluation was also performed

in terms of the complication-free tumour control probability, P+. The physical and

radiobiological evaluation corresponding to plans obtained by the inverse planning

simulated annealing algorithm (IPSA) and the hybrid inverse planning optimization

(HIPO) have been compared with corresponding ones for plans obtained using a manual

graphical optimization method.

The main observations of this work are that well tuned class solutions of inverse

optimization methods are able to produce similar dose volume histograms to those

produced with manual graphical optimization and inverse methods have the potential

to spare organs at risk while delivering acceptable dose to the target. In addition, radiobiological

indexes such as the P+ can be useful complements to physical parameters

in treatment plan evaluation.

Place, publisher, year, edition, pages
Stockholms universitets förlag, 2015
Keywords
Brachytherapy, treatment planning optimization, IPSA, HIPO, radiobiology, complicationfree
National Category
Physical Sciences
Research subject
Medical Radiation Physics
Identifiers
urn:nbn:se:su:diva-121548 (URN)
Opponent
Supervisors
Available from: 2016-05-03 Created: 2015-10-07 Last updated: 2022-02-17Bibliographically approved
2. Optimization of brachytherapy treatment planning and radiation field studies in the vicinity of the Leksell Gamma Knife®
Open this publication in new window or tab >>Optimization of brachytherapy treatment planning and radiation field studies in the vicinity of the Leksell Gamma Knife®
2024 (English)Doctoral thesis, comprehensive summary (Other academic)
Abstract [en]

Carcinoma of the cervix is a global problem, and brachytherapy is one of the main radiation therapy components used in the treatment of patients with this type of cancer. With the advent of scientific and technological developments in treatment planning, inverse optimization in brachytherapy and its thorough comparison with traditional manual optimization methods is warranted. In this work, the physical parameters; minimum dose received by 98% and 90% of the target volume represented by D98 and D90, respectively, were used to evaluate the treatment plans with respect to the target while the minimum dose received by 2cm3 volume, D2cm3, was used to investigate complications in organs at risk. The conformity index, COIN, was used to describe the coverage of the target by the prescribed dose and the fraction of each organ at risk volume that receives the critical dose, which may cause complications. The treatment plan evaluation was also performed in terms of the radiobiological parameter complication-free tumor control probability, P+. The physical and radiobiological evaluation corresponding to two brachytherapy inverse planning algorithms have been compared with homologous manual graphical optimization plans. The main observations from this study are that well tuned class solutions of inverse optimization methods may produce similar dose volume histograms to those produced with manual graphical optimization plans, and inverse methods have the potential to spare organs at risk while delivering an acceptable dose to the target. In addition, radiobiological indexes such as the P+ can be useful complements to the physical parameters in treatment plan evaluation. The Elekta Leksell Gamma Knife® unit has been successfully utilized in the management of intracranial malignancies for more than half a century. As required by national and international regulation, adequate knowledge of risks posed by ionizing radiation instrumentation is necessary in order to protect the patient, workers, public, and the environment. In that perspective, the Nuclear Physics research group at the Department of Physics of Stockholm University (Stockholm, Sweden) in collaboration with Elekta Instrument AB (Stockholm, Sweden) have conducted investigation on the radiation field in the vicinity of the Gamma Knife utilizing the High Purity Germanium (HPGe) detector. As part of the ongoing research, the principal objective of the present work was to improve the modeling and characterization of the radiation field around the Gamma Knife to interrogate the efficacy of the National Council on Radiation Protection and Measurements (NCRP) methodology for structural shielding design and evaluation for the Leksell Gamma Knife treatment room. Acquisition of high-resolution γ-ray spectra and ambient dose equivalent H*(10) in the field of the Gamma Knife-PerfexionTM took place at the Division of Neurosurgery, Radiumhemmet (Department of Oncology), Karolinska University Hospital (Stockholm, Sweden). A p-type coaxial HPGe detector and a satellite survey meter were utilized to acquire the γ-ray spectra and H*(10), respectively. The measured configurations were simulated on Pegasos Monte Carlo system. A phase space on a cylindrical surface enclosing the Gamma Knife with open doors and the phantom assembled was used as the source of radiation. About 4·107 γ photons were collected on the phase space corresponding to 2·1012 decays. With the Gamma Knife doors open, most of the radiation was measured in the forward direction up to θ = 45o in relation to the z axis. The Monte Carlo simulations reproduced the measured results and; therefore, good agreement was achieved between response measurements and simulated spectra. The recent Gamma Knife models PerfexionTM, IconTM, and EspritTM are dosimetricaly equivalent and represent more than half the total number of Gamma Knife systems installed around the globe. Given the increasing importance of these machines in the treatment of patients with intracranial malignancies, research and development of tools for radiological protection around the Gamma Knife unit is topical and will continue to evolve, especially with the advent of new models or versions of Gamma Knife units integrating novel technology such as the satellite cone-beam computed tomography (CBCT) imaging system. The Monte Carlo method is a powerful tool for simulation of radiation transport in matter. It reproduced the measured results and may be used in structural shielding design calculations for γ-photons from the Gamma Knife.

Place, publisher, year, edition, pages
Stockholm: Department of Physics, Stockholms universitet, 2024. p. 108
Keywords
Brachytherapy, treatment planning optimization, radiobiology, complication-free tumor control probability, Leksell Gamma Knife, stereotactic radiation surgery, HPGe detector, detector response, Monte Carlo simulation, dosimetry, radiological protection
National Category
Physical Sciences
Research subject
Medical Radiation Physics
Identifiers
urn:nbn:se:su:diva-233393 (URN)978-91-8014-933-4 (ISBN)978-91-8014-934-1 (ISBN)
Public defence
2024-10-25, Room FB42, AlbaNova Campus, Roslagstullsbacken 21, Stockholm, 10:00 (English)
Opponent
Supervisors
Available from: 2024-10-02 Created: 2024-09-11 Last updated: 2024-10-08

Open Access in DiVA

No full text in DiVA

PubMedScopus

Authority records

dos Santos Matias, LucílioToma-Dasu, Iuliana

Search in DiVA

By author/editor
dos Santos Matias, LucílioToma-Dasu, Iuliana
By organisation
Department of Physics
In the same journal
Anticancer Research
Cancer and Oncology

Search outside of DiVA

GoogleGoogle Scholar

pubmed
urn-nbn

Altmetric score

pubmed
urn-nbn
Total: 165 hits
CiteExportLink to record
Permanent link

Direct link
Cite
Citation style
  • apa
  • ieee
  • modern-language-association-8th-edition
  • vancouver
  • Other style
More styles
Language
  • de-DE
  • en-GB
  • en-US
  • fi-FI
  • nn-NO
  • nn-NB
  • sv-SE
  • Other locale
More languages
Output format
  • html
  • text
  • asciidoc
  • rtf